GBT Troubleshooting

Focusser won’t Focus

Warning

TA or OS Only should physically handle the focusser

Note

This is very problematic, there are various strategies.

Note

Strategy 1: Resetting the Focusser

  • Close FocusMax and CCD Soft on the Computer

  • At the telescope, check the focusser is unlocked

  • Use the focus hand control to drive to “physical zero”

    • i.e. there is no gap where the focusser moves outwards

    • ignore whatever value is on the hand controller

  • Unplug focusser - breathe - plug back in

  • Check focusser control now reads 0000

  • Re-open FocusMax

    • Focus should move to last set value

    • Check value in FocusMax

    • Check value matches on manual control

  • Re-try Focus

See also